PRODUCTS
BM-504 Patient Monitor
|
|
|
|
Feature set
- Compact, light, easy for carrying and handling;
- Large, bright, easy-to-read LED display of measurement;
- Unique character and icon display of information;
- Storage and review of NIBP and SpO2 data up to 800 groups;
- Built-in rechargeable Li-battery for 10 working hours;
- Low battery power indicator;
- Automatic power-off;
- External printer available
|
|
Technical Specification |
|
 |
|
| NIBP |
• Measurement Method: Oscillometric
• Patient Range: Adult and pediatric
• Measurement Unit: mmHg/kPa selectable
• Measurement Range:
• adult (30 - 255 mmHg)
• pediatric (20 - 160 mmHg)
• Measurement Mode: manual, auto, STAT
• Measurement Type: SYS, DIA, MAP
• Meas. Interval in AUTO mode: 1, 2, 3, 5, 10, 15, 20, 30, 60 minutes
• Resolution: 1 mmHg
• Over-pressure Protection:
• adult (300 mmHg)
• pediatric (220 mmHg) |
| Oxygen Saturation (SpO2) |
• Measurement Range: 35 - 100%
• Accuracy: 2%(70-100%), 0-69% unspecified
• Resolution: 1% |
| Pulse Rate |
• Measurement Range: 0-250 bpm
• Accuracy: 2bpm
• Resolution: 1 bpm |
| Alarm |
• Mode: audio, visual
• Type: SYS, DIA, MAP, PR, SpO2 |
| Storage and Review |
800 groups of NIBP, SpO2, PR data with corresponding date, time, and number |
| Display |
• Parameter Display: 7-segment LEDs
• Information Display: fixed characters and icons lit by single-color LEDs
• Power Display: dual-color LED segments |
| External Printer |
• Model: Thermal microprinter
• Interface: TTL standard serial interface |
| Thermal Recording Paper |
• Width: 57 +0/-0.5mm
• Thickness: 65um
• Outer Diameter of Paper Roll: <+30mm
• Accuracy: ± 2%
• Resolution: 1 bpm |
| Environment |
• Temperature (Operation): 0-45゚C
• Humidity (Operation): 30-85% (non-condensing)
• Temperature (Storage): -25-85゚C
• Humidity (Storage): 30-95% (non-condensing) |
| Power Requirements |
• Type: Lithium-ion rechargeable (internal) battery
• Operating Time: 10 hours11 |
|
|
|
|
|